Rigol DM3068 Digital-Multimeter, DMM, DM3068 数据表

产品代码
DM3068
下载
页码 171
RIGOL 
User’s Guide for DM3068 
 
1-9 
Rear Panel 
 
 
 
 
   
 
 
Figure 1-6 Rear Panel Overview 
 
1.  Current Input Fuse 
The multimeter uses two kinds of fuses for input protections of small and large 
current ranges. The internal fuse of the multimeter provides 10 A maximum 
input protection for the large current range and when the input current exceeds 
13.5 A, the fuse will blow out; while the current input fuse at the rear panel of 
the instrument provides 500 mA maximum input protection for the small current 
range and when the input current exceeds 650 mA, the fuse will blow out. The 
large current input fuse has been installed when the instrument left factory. To 
replace the small current fuse, please: 
 
1)  Turn off the multimeter and remove the power 
cord. 
2)  Turn the fuse seat counterclockwise as shown in the 
figure using a straight screw driver and then pull 
out the fuse seat.   
                                     
1.Current Input Fuse   2. LAN          3.GPIB            4.Power Socket 
6.Ext Trig                      8.RS232       10.Voltage Selector 
5.VMC                7.USB Device           9.Power Fuse        11.Power Switch