User’s Guide for DM3068 
 
2-5 
Basic Measurement Functions 
 
To Measure DC Voltage 
 
Range: 200 mV, 2 V, 20 V, 200 V, 1000 V 
Max Resolution: 100 nV (in the range of 200 mV) 
Input Protection: a 1000 V protection is available in all ranges and a 10% 
overrange for all ranges except 1000 V range. If the reading exceeds 1050 V in 1000 
V range, “OVER LOAD” will be displayed.   
 
Operating Steps: 
 
1.  Enable the DCV measurement   
Press 
  at the front panel to enter the following interface. By default, DCV 
measurement is selected at each power-on. 
   
 
2.  Make connection     
Connect the test leads with the measured signal by referring to 
Measurement Connections
 
3.  Set the range and resolution 
Specify an appropriate range and resolution according to the measured signal 
and measurement requirements. A high voltage icon “ ” will appear at the 
upper left corner of the screen when the 1000 V range is selected.   
   
4.  Set the measurement parameters (Optional) 
The DCV measurement allows users to configure the integration time, dc 
impedance and auto zero. For details, please refer to Measurement 
Configuration
.