2-11 
To Measure AC Current 
 
Range: 200 μA, 2 mA, 20 mA, 200 mA, 2 A, 10 A 
Max Resolution: 0.1 nA (in 200 μA range) 
Input Protection: the multimeter uses two kinds of fuses for input protection: the 
small current input fuse (500 mA high speed fuse) at the rear panel and the built-in 
large current input fuse (10 A). 10% overrange is available on all ranges except 10 
A. If the reading exceeds 10.5 A in 10 A range, “OVER LOAD” will be displayed. 
 
In order to obtain more accurate measurement results, DM3068 deals with small 
and large input currents separately. The multimeter will use small current 
measurement mode for current within 200 mA and use large current measurement 
mode for current that is equal to or higher than 2 A. 
 
Operating Steps:   
 
1.  Enable the ACI measurement 
Press 
  at the front panel to enter the following interface. 
 
 
2.  Make connection 
Connect the test lead with the measured signal by referring to 
Measurement Connections.   
 
3.  Set the range   
Specify an appropriate range (the resolution is fixed at 6
1
/
2
 digits) according to 
the measured signal and measurement requirements. A high voltage icon “ ” 
will be displayed at the upper left corner of the screen when the 10 A range is 
selected. 
 
4.  Set the measurement parameters (Optional) 
The DCI measurement allows users to configure the filter. For details, please 
refer to Measurement Configuration