2)  Only use 10 A and LO terminals for measurements when the measured 
current AC+DC RMS value goes within 200 mA and 10 A. 
3)  Select a proper current input terminal according to the estimated current 
magnitude before connect the multimeter to AC supplies if you want to 
use current measurement. 
4)  The current into 10 A cannot exceed 13.5 A, otherwise it will blow out 
the internal fuse; while the current into the LO Sense/200 mA terminal 
cannot exceed 650 mA, otherwise the current fuse from the rear panel 
may be blown out.     
 
IEC II Overvoltage Protection.
 
In order to prevent electric shock, DM3068 provides overvoltage protection for 
line-voltage mains connections meeting both of the following conditions: 
1  The HI and LO input terminals are connected to the mains under 
Measurement Category II conditions, defined below. 
2  The mains are limited to a maximum line voltage of 300 VAC. 
 
Warning: IEC II includes electrical devices connected to mains at an outlet on a 
branch circuit. Such devices include most small appliances, test equipments and 
other devices that inserted into a branch socket.   
DM3068 may be used to make measurements with the HI and LO inputs connected 
to mains in such devices (up to 300 VAC), or to the branch socket itself. However, 
DM3068 may be used with its HI and LO inputs connected to mains from neither 
permanently installed electrical device such as a main circuit-breaker panel, 
sub-panel disconnected box nor wired motors. Such devices and circuits are readily 
to beyond the protection from DM3068. 
 
Note: voltages above 300 VAC may be measured only in circuits that are isolated 
from mains. However, a transient overvoltage is also present in such circuits. 
DM3068 was designed to safely withstand occasional transient overvoltage up to 
2500 Vpk. Do not use this device to measure circuits whose transient overvoltage 
may exceed this level.