Rigol DM3068 Digital-Multimeter, DMM, DM3068 数据表

产品代码
DM3068
下载
页码 171
RIGOL 
User’s Guide for DM3068 
2-34 
TC 
 
The TC (thermocouple) is one of the most commonly used industrial temperature 
detection device. It converts the temperature value into voltage reading, and has a 
very large temperature measurement range. 
 
The common thermocouples are B (Pt Rh 30- Pt Rh6), E (NI CR-WRCK), J 
(Fe-WRCK), K (NI CR-NiSi), N (NiCrNi-NiSi), R (Pt Rh13 -Pt), S (Pt Rh10-Pt) and T 
(Cu-WRCK). Among them, B, R and S are precious metal thermocouple, while E, J, 
K, N and T are cheap metal thermocouples. For the reference table about each 
thermocouple, please see the ITS-90 standards.   
 
The DM3068 provides two cold junction compensation modes: internal cold 
junction temperature
 & user-defined cold junction temperature
 
Internal cold junction temperature refers to the internal temperature of the 
banana jack detected by the multimeter. Connect the thermocouple wire to the 
inside of the banana jack and prevent air flow from interfering the temperature 
around the banana jack. The radiating structure on the banana jack connected 
with the thermocouple wire would cause additional cold junction temperature error. 
About 3-minute warm-up is needed to build up the thermal balance after the cold 
junction is connected. 
 
User-defined cold junction temperature refers to the temperature at the cold 
junction compensation point of the thermocouple. The internal temperature of the 
banana jack might be different from the actual temperature at the cold junction 
compensation point. If external cold junction compensation equipment is used, the 
thermocouple voltage signal compensated can not be connected to the banana 
jack through the compensation lead; otherwise, additional cold junction error will 
be caused. The accuracy of the measurement when using simulated cold junction 
depends on the accuracy of cold junction compensation. 
 
To use the TC sensor, you should select the type of the thermocouple and the 
compensation mode of the cold junction. 
Press TC  Type to select the desired thermocouple type and press 
Press TC  JUNC, select INT or SIM to input correct cold junction temperature 
(with -273 and 999) and then press