Rigol DM3068 Digital-Multimeter, DMM, DM3068 数据表

产品代码
DM3068
下载
页码 171
RIGOL 
User’s Guide for DM3068 
2-40 
Measurement Configuration   
 
Most measurement parameters are user-defined. Changing a measurement 
parameter will change the measurement precision and speed as well as the input 
impedance. An appropriate measurement parameter based on the actual 
application will ensure faster measurement or higher measurement precision. 
 
The default measurement configurations of the multimeter can ensure the 
accuracy of the measurement results in most cases. Users can directly use these 
defaults for any measurement or modify the parameters of the measurement 
function as required. 
 
The parameters for different measurement function differ, see table below. 
 
Table 2-5 Measurement parameter 
Functions   
Parameters 
DCV 
integration time, DC impedance, Auto zero (AZ) 
ACV 
AC filter 
DCI 
integration time, Auto zero (AZ) 
ACI 
AC filter 
OHM (2WR, 4WR) 
Integration  time,  Auto zero  (AZ),  Offset compensation 
(OC) 
CAP 
N/A 
CONT 
Short-circuit resistance 
DIODE 
Current Source 
FREQ/PREIOD 
Gate time, AC filter 
SENSOR 
N/A 
 
To set the measurement parameter, select a measurement function and then press 
. For more details, please refer to the following content.