User’s Guide for DM3068 
 
2-47 
Current Source 
 
Current source applies to the diode measurement function. The current selected is 
applied to the diode under test and the multimeter measures the voltage drop 
between the two ends of the diode to determine the polarity of the diode. 
 
Note: when measuring diode with relatively larger drain current, do not use the 
small current source (as the forward voltage drop measured might be too low to 
determine the polarity of the diode). 
 
In DIODE measurement, press 
  to enter the following interface. The current 
sources available are 1 mA, 100 μA, 10 μA, 2 μA and 200 nA and the corresponding 
voltage ranges are -0.2 V ~2.2 V, -0.2 V ~4.2 V, -0.2 V ~4.8 V, -0.2 V ~5.2 V and 
-0.2 V ~5.5 V. You can also press 
  to select the desired current source.