Gw Instek GDM-8342USB Digital-Multimeter, DMM, GDM-8342USB 数据表
产品代码
GDM-8342USB
GDM-834X Series User Manual
72
Math Measurement
Math Measurement Overview
Math measurement runs three types of mathematical operations,
MX+B, 1/X and Percentage based on the other measurement
results.
MX+B, 1/X and Percentage based on the other measurement
results.
Applicable
Measurements
Measurements
The math function can be used with the
following basic measurement functions:
ACV, DCV, ACI, DCI, Ω, Hz/P, TEMP
following basic measurement functions:
ACV, DCV, ACI, DCI, Ω, Hz/P, TEMP
Overview of Math
Functions
Functions
MX+B
Multiplies the reading (X) by the
factor (M) and adds/subtracts
offset (B).
factor (M) and adds/subtracts
offset (B).
1/X
Inverse. Divides 1 by the reading
(X).
(X).
Percentage
Runs the following equation:
100%
Reference
Reference)
-
(ReadingX
x
Measure MX+B
Procedure
1. Press SHIFT → MX/MN to enter the MATH
menu.
The MX+B setting appears. The M factor will be
flashing, indicating that the M factor is to be set.
flashing, indicating that the M factor is to be set.
2. Use the Left and Right arrow keys to navigate
to the digit to be edited or to select the decimal
point.
point.
Use the Up and Down arrow keys to edit the
selected digit or to place the position of the
selected digit or to place the position of the