2.  From the front panel, use a screwdriver to rotate the fuse carrier several turns counter-clockwise. 
Take the fuse carrier out of the socket. 
3.  Remove the fuse and replace it with the same type (T1AL, 250V, 5×20mm). 
   
CAUTION:  Do not use a fuse with a higher current rating than specified or instrument may be 
damaged. If the instrument repeatedly blows fuses, try to find out the reason before 
replacing the fuse. 
 
4.  Install the new fuse by reversing the procedure above. 
3.4    Measuring Resistance   
Model 2831E Resistance measurement range: 200 
Ω, 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20 MΩ; Maximum 
resolution: 10 
mΩ (on 200 Ω range) 
Model 5491B Resistance measurement range: 500 
Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 50 MΩ; Maximum 
resolution: 10 
mΩ (on 500 Ω range) 
 
3.4.1    Connections 
Assuming  the multimeter  is under factory default conditions, the basic procedure for measuring 
resistance is as follows: 
1.  Connect the test leads between 
and COM. 
2.  Select resistance measurement function by pressing 
 
3.  Press 
  toggles auto ranging. Notice the AUTO annunciator is displayed with auto ranging. If 
you want manual ranging, use the RANGE 
  and 
  keys to select a measurement 
range consistent with expected resistance. 
4.  Connect test leads to the resistance as shown in Figure 3-3: 
 
CAUTION:  Do not apply more than 1000  V peak
 
between
 
  and  COM  or instrument 
damage may occur.   
 
5.  In manual range, if the “OVL.D” message is displayed, press up 
  key to select a higher range 
until a normal reading is displayed (or press 
  key for auto ranging). Use the lowest possible 
range for the best resolution.   
6.  Take readings from the display.