5 Performing the measurement 
14 
5
 Performing the measurement 
 
CAUTION 
High thermal radiation (e.g. due to sun, fire, furnaces) 
Detector may be damaged! 
-
  Do not point the camera at objects with temperatures exceeding 
500 °C. 
 
 
 
Ideal framework conditions 
 
  Building thermography, investigating the building shell:  
Considerable temperature difference between inside and outside required 
(ideal: ≥ 15 °C / ≥ 27 °F).  
 
  Consistent weather conditions, no intensive sunlight, no precipitation, no 
strong wind. 
 
  To ensure maximum accuracy, the camera requires an adjustment time of 
10 minutes after it is switched on. 
 
Important camera settings 
 
  Emissivity and reflected temperature must be set correctly in order to 
precisely determine the temperature. Subsequent adjustment is possible via 
the PC software, if required. 
 
  When auto-scaling is activated, the colour scale is continuously adjusted to 
the min./max. values of the current measurement image. This means that 
the colour assigned to a specific temperature is changing constantly! To be 
able to compare multiple images based on the assigned colour, scaling 
must be set manually to fixed values, or must be subsequently adjusted to 
uniform values using the PC software. 
 
5.1
  Saving an image 
 
- Press 
trigger
 
 
The image is automatically saved. 
 
testo 870-2: regardless of the image 
type set, an infrared image is saved 
with an attached real image.