Testo 625 Thermo-Hygrometer 0563 6251 用户手册
产品代码
0563 6251
5. Operation 29
1
Select the desired option with
and confirm with
:
· YES: Switches automatic probe detection on
(recommended).
· no: Switches automatic probe detection off.
no
was selected:
2
Use
/
to set the probe ID manually and
confirm with
.
º
Continue with objective
T
O SET
A
UTO
O
FF
.
YES
was selected:
- Automatic probe detection is started.
- Once a radio probe is found, the probe ID is
- Once a radio probe is found, the probe ID is
displayed. If no probe is found, NONE lights up.
If more than one radio probe is switched on:
i
If necessary, select the desired probe with
/
.
Possible causes why probes are not found:
· The radio probe is not switched on or the
battery of the radio probe is spent.
· The radio probe is outside the range of the
measuring instrument.
· Sources of interference are influencing the radio
transmission (e.g. reinforced concrete, metal
objects, walls or other barriers between
transmitter and receiver, other transmitters of the
same frequency, strong electromagnetic fields).
objects, walls or other barriers between
transmitter and receiver, other transmitters of the
same frequency, strong electromagnetic fields).
i
If necessary, rectify the possible causes for the
disruption to the radio transmission and start
configuration mode again.
disruption to the radio transmission and start
configuration mode again.
2
Press
to change to the next function.