Calibration
Adjustment Procedures
DAQ on- board 5 V calibration
If in the performance verification procedure is not accurate,
adjustment procedure is required. The following flowchart
shows the steps for 5 V calibration.
adjustment procedure is required. The following flowchart
shows the steps for 5 V calibration.
Figure 2-14 The flowchart showing the steps for 5 V calibration
Step 2: 5 V calibration
Connect the DAQ device to the computer with USB cable.
Monitor the DMM readings.
Tune the 5 V until you get the voltage in between 4.9999 to 5.0001 V.
Send “
Connect the DAQ device to the computer with USB cable.
Monitor the DMM readings.
Tune the 5 V until you get the voltage in between 4.9999 to 5.0001 V.
Send “
CAL:VREF:TUNE MINUS” (to decrease voltage),
Send “
CAL:VREF:TUNE PLUS” (to increase voltage),
Send “
CAL:VREF” (to save to EPROM)
Step 1:
Open the DAQ device metal casing.
Power on DAQ device and warm up for 30 minutes.
Connect the DMM to U37 on the daughter board.
Open the DAQ device metal casing.
Power on DAQ device and warm up for 30 minutes.
Connect the DMM to U37 on the daughter board.
Step 3:
After save to EPROM,
power cycle the unit.
After save to EPROM,
power cycle the unit.
Step 4:
Reconfirm the 5 V
reading.
Reconfirm the 5 V
reading.