Gigahertz Solutions MK70-3D -Analyser, Electric smog meter, 930-243 数据表

产品代码
930-243
下载
页码 32
350-027_NFA1000400-6-2_EN-von-susanne as of October 2011
- 18 -
© Gigahertz Solutions GmbH
6 Power Supply
6.1 Internal Rechargeable Battery
The instrument features high performance Lithium Ion rechargeable batteries
(2 x 3.7 V). When fully charged, they will last for up to 36 hours of continuous data
logging
13
.
6.2 Charging
To charge the battery set the AC adapter switch to 15 Volts. If the device is
switched off while charging, the display will show “CHAr” and the potential runtime
of the instrument with the current capacity status in an infinite loop. Measurements
or recordings during the charging process are technically possible, but not speci-
fied, and therefore not recommendable.
With a completely discharged battery only about one hour of charging time is suffi-
cient to operate the instrument for a couple of hours. To achieve the maximum op-
eration time, charging will take 12 to 15 hours. Once the full capacity is reached, the
instrument will stop charging automatically and will show "FULL" on the LCD.
6.3 External Power Supply
Switch the AC adaptor to 12 Volts. With a 32 GB SD card and an external power
supply (AC adaptor) you can continuously log at full speed for over two months.
Longterm recordings of the 3D magnetic field and the AC entry can also be done
using external power from an AC adapter. In this case, the adapter should be placed
as far as possible from the instrument and turned in such a way that the resulting
magnetic field is as low as possible, as it can influence the readings.
14
13
Please note that Li-Ion batteries, even when treated with care, will lose some of their original capacity within
one or two years and their typical life expectancy is 250-500 complete charge cycles. In this case please always
replace both batteries. For initialisation of the indication of the remaining battery life of the instrument please fully
unload and recharge the instrument once. If possible store in a cool place..
14
To determine the emission created by the AC adapter at the location of the NFA, proceed as follows:
-
Place the instrument and adaptor into their measurement location
-
Plug the AC adapter into the wall and to the NFA. Wait until the readings on the display stabilize.
-
Disconnect the adapter from the NFA without changing its position and orientation; wait for the display to
stabilize.
This procedure will also be logged, and should be marked with an audio note. This will be accessible when
evaluating the logged data on the PC. Now potential disturbances can be noted on each axis and corrections
be made, if necessary.