Extech Layer-thickness tester, paint-coat measurement CG204 数据表
产品代码
CG204
CG204-EU-EN V2.1 3/11
4
Programming Menu
The meter can be configured and calibrated through simple button presses in the
programming menu. Press the MENU button to access the menu and refer to the menu
‘tree’ below. The menu uses UP/DOWN, SELECT, BACK, & ESC presses for navigation
and selection. In the table below the factory default settings are in bold with an asterisk.
Each parameter is explained in detail in the subsequent sections.
programming menu. Press the MENU button to access the menu and refer to the menu
‘tree’ below. The menu uses UP/DOWN, SELECT, BACK, & ESC presses for navigation
and selection. In the table below the factory default settings are in bold with an asterisk.
Each parameter is explained in detail in the subsequent sections.
Top level
Sub level 1
Sub level 2
Notes
STATISTICAL VIEW
AVG*
Average of a series of readings
MIN
Lowest of a series of readings
MAX
Highest of a series of readings
NO
Number of sampled readings
SDEV
Standard deviation of a series
OPTIONS
Measure mode
Single*
One reading at a time
Continuous
Continuous measurements
Working mode
Direct*
Readings are not stored in groups
Group 1…4
Store readings in groups
Probe used
Auto*
Meter automatically selects mode
Fe
Ferrous measurement mode
No Fe
Non-ferrous measurement mode
Unit settings
µ
m*
Micrometers
mils
Mils = mm * 2.54 / 100
mm
Millimeters
Backlight
ON*
Enables backlight operation
OFF
Disables backlight operation
LCD Statistics
AVG*
Average of a series of readings
MAX
Lowest of a series of readings
MIN
Highest of a series of readings
SDEV
Standard deviation of a series
Auto Power OFF
Enable*
Allows auto power off to operate
Disable
Defeats the auto power off feature
LIMIT
Limit settings
High Limit
High Alarm alerts user when reached
Low Limit
Low Alarm alerts user when reached
Delete Limits
Clear the alarm limit values
DELETE
Current Data
Delete current data
All Data
Delete all stored data
Group Data
Delete data stored data plus alarm
and calibration data
and calibration data
MEASUREMENT
VIEW
View stored data in all groups
CALIBRATION
Enable
Allow calibration access
Disable
Lock out the calibration mode
Delete Zero N
Clear the zero cal. data (non-ferrous)
Delete Zero F
Clear the zero cal. data (ferrous)
NOTE: Disable the Auto Power OFF feature before lengthy programming to avoid
inconvenient automatic power down while programming.
inconvenient automatic power down while programming.