The memory function 
The  SC  6400  can  store  up  to  85 
measurements  in  the  memory  area. 
Together 
with 
each 
measurement 
(systolic,  diastolic  and  heart  rate),  the 
date  and  time  are  also  saved.  The 
measurement  with  the  longest  residence 
time  in  the  memory  is  assigned  number 
1.  All  other  measurements  increment  by 
one until number 85. Then the memory is 
full.  The  next  measurement  occupies 
memory  location  85.  The  measurement 
formerly  at  location  85  changes  to 
location  84  (etc.,  83
82  ....  2
1).  The 
oldest  measurement  formerly  at  location 
1 is lost. 
 
Memory recall 
With  the  MEMORY  button  you  can  recall 
the memories.  
 
Delete a blood pressure value from the 
memory 
Press  the  MEMORY  button  to  view  the 
blood  pressure  value  you  want  to  delete. 
Pressing  the  0/I-START  button  instantly 
(and  holding  it  for  at  least  4  secs.) 
displays:  
dEL (for "delete") and the number of 
the memory location. 
Press the 0/I-START button instantly and 
the value will be deleted. 
 
Delete all blood pressure values in the 
memory 
Press the MEMORY button to display any 
blood pressure value. Pressing the 
POWRER key instantly (and holding it for 
at least 4 secs.) displays: 
dEL (for "delete") and the number of 
the memory location 
Press the MEMORY button instantly and 
the LCD displays: 
dEL (for "delete") and ALL (for all 
memory locations). 
Press the 0/I-START button instantly and
 
- 6 - 
all  values  will  be  deleted.  Time  and  date 
are not deleted. 
 
WHO*-Interpretation of blood pressure 
readings                                    
 
29.12.2007
   
 
Meaning 
Systolic / 
mmHg 
Diastolic / 
mmHg 
optimal 
< 120 
< 80 
normal 
< 130 
< 85 
high-normal 
130 - 139 
85 - 89 
mildy 
hypertonic 
140 - 159 
90 - 99 
hypertonic, 
medium 
severity 
160 - 179 
100 - 109 
severly 
hypertonic 
> 180 
> 110 
   
The  World  Health  Organization  (WHO) 
recommends  medical  treatment,  when 
the blood pressure value exceeds 140/90 
mmHg.  
 
Examples: 
 
 
 
 
Cleaning and maintenance  
  Do  not  expose  the  BP  monitor  to 
extreme  temperature,  moisture,  dust  or 
direct sunlight. 
  The  cuff  contains  a  sensitive  airtight 
bulb.  Treat  it  with  care  and  avoid  any 
stress by torsion or kinking. 
  Clean  the  BP  monitor  with  a  soft,  dry 
cloth.  Do  not  use  fuel,  thinner  or  other 
solvent.  Remove  specks  on  the  cuff 
carefully with a wet cloth and soap water.  
Do not wash the cuff. 
  Do  not  drop  the  BP  monitor  or  treat  it 
with force.  
 
- 7 - 
 Do not try to open the BP monitor. This 
will make the factory calibration invalid
 
 
Troubleshooting and remedies 
If 
fault 
is 
detected 
during 
measurement,  the  BM  monitor  displays 
„Err“  and  a  symbol  (code  for  service 
personnel).  
From many years of experience we know 
that  most  faults  are  the  consequence  of 
wrong  use.  The  most  frequent  causes  of 
faults are excessive movement during the 
measurement,  wrong  application  of  the 
cuff,  use  of  wrong  batteries  or  use  of  a 
rechargeable battery. 
 
 
 
Message 
 
 
Error description 
 
Correction 
 
 
The pressure 
measured was 
lower than 
20mmHg. 
Please measure 
again. 
 
 
 
The pressure 
measured was 
higher than 
300mmHg.  
Please measure 
again. 
 
 
 
 
Pumping error. 
 
Please check cuff 
and try again. 
 
 
The pressure can 
not be measured 
due to signal 
noise. 
Please measure 
again. 
 
 
 
 
The pumping 
pressure is 
higher than 
300mmHg. 
Please measure 
again. 
 
or no  
display 
 
Low batteries. 
Please check and 
replace the batteries 
if necessary. 
 
Wait  at  least  5  minutes  before  repeating 
the measurement. 
 
Periodic re-calibration 
We  recommend  periodic  checks  of  the 
pressure indication every 2 years. 
 
Referenced norms 
For  registration,  a  defined  number  of 
blood pressure monitors was subjected to 
stringent clinical testing. These monitors  
 
- 8 - 
used  the  same  program  for  blood 
pressure  measurement  as  your  monitor 
uses. Thus, this program is also regarded 
as clinically tested. 
The  manufacture  of  the  blood  pressure 
monitor  complies  with  the  European 
norms  for  blood  pressure  monitors:  EN 
60601-1, EN 1060-1 to EN 1060-3. 
 
Electromagnetic compatibility:  
The blood pressure monitor complies with 
the  requirements  of  European  norm  EN 
60601-1-2.  
The 
requirements 
of 
EU 
directive 
93/42/EEC  concerning  medical  devices, 
class IIa are complied with. 
 
Warranty 
We  grant  a  warranty  of  two  years  for  the  
blood  pressure  monitor  SC  6400.  Within 
this period, defects of the BP monitor will 
be  repaired  free  of  charge.  You  can  also 
have  your  BP  monitor  repaired  by  our 
customer  service  after  the  warranty 
period.  We  will  provide  you  with  a  free 
cost estimate for the repair.   
The warranty does not cover damage due 
to  improper  handling,  accident,  natural 
wear and tear (cuff), non-compliance with 
the  instructions  for  use  or  unauthorized 
modifications of the monitor. 
Excluded from the warranty are:  
the 
packaging, 
batteries 
and 
the 
instructions for use. 
 
   
Disposal note  
The  disposal  of  batteries  is  regulated. 
Spent batteries are collected at collection  
points  provided  by  battery  dealers  and 
the communities. 
Dispose of the blood pressure monitor as 
prescribed.  i.e.,  at  a  collection  point  for 
electrical scrap. 
 
 
 
- 9 - 
 
Specifications                              
Measuring method:  Oscillometric  
Measuring range:     Pressure: 20 to 300mmHg 
                          
Heart rate: 40 to 200/min 
Measuring accuracy: stat. pressure: +/- 3mmHg                         
Heart rate: +/- 5 % 
Measuring value memory:   
        
memory with 85 locations  
for systolic / diastolic / 
heart rate, date and time  
Display:                    LC display 
Automatic switch-off:   
            
after approx. 1 minute of  
 
inactivity  
Voltage supply:   
2 x LR03 (alkali batteries), 
Working temperature range: 
                                 +10°C  -  +40°C 
Storage temperature range:   
                                 - 10°C  -  +50°C 
Weight (without batteries):     approx. 100 g  
Cuff length:   
for wrist circumference  
from 135mm to 195mm        
 
Specifications may change as technical 
improvements take place
 
 
*WHO     –  World Health Organization  
 
 
    
      Service client 
SCALA Electronic GmbH         SES GmbH  
Ruhlsdorfer Str.95  
      Ruhlsdorfer Straße 95 
14532 Stahnsdorf                  14532 Stahnsdorf 
            
Explanation of symbols 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Rev. 25/07/2012/GB 
 
 
 
- 10 -