Tdk Lambda DTM65PW240C 65W Medical AC-DC External Power Supply 24Vdc 2700mA DTM-65PW-240-C 数据表

产品代码
DTM-65PW-240-C
下载
页码 4
DTM65-C Series
1
DTM65-C Series
40W to 65W Medical  AC-DC External Power Supplies
Key Market Segments & Applications
Medical, Industrial Computing,
Life Sciences / Laboratory
• ErP
1,
CEC
2
and EISA
3
Compliant Models
• Wide Range AC Input
• Medical Certifications
DTM65-C Features and Benefits
Features
Benefits
• CEC and EISA Compliant
• Easier system compliance
• Wide Range AC Input
• Global Operation
• 87% Average Efficiency
• Consumes less energy
Specifications
ITEMS
DTM65-C
Input Voltage range
90 - 264VAC (47 - 63Hz)
Inrush Current
A
100A at 230VAC input, 25°C ambient cold start
Input Current
A
2A at 90VAC, Full Load
Power Factor
None
Leakage Current (264VAC 63Hz)
mA
<300uA
Hold Up Time (Typ)
ms
16ms at 230VAC input
Temperature Coefficient
±0.05%/°C
Voltage Accuracy
%
±1%
Adjustment Range
None
Minimum Load
A
None
Total Regulation
12V Model: ±5%, >15V models ±3%
Ripple & Noise
%
1% or 100mV whichever is the greater
Short Circuit Protection
-
Continuous - hiccup mode
Overvoltage Protection
V
110 - 150% of nominal (Cycle input power to reset)
Average Efficiency
%
87%
Operating Temperature
°C
0 to +50°C
Storage Temperature
°C
-10 to +70°C
Humidity (non condensing)
10 - 90% RH
Cooling
-
Convection
Withstand Voltage
Input to Ground 1.5kVAC, Input to Output 4kVAC (Output is connected to Ground)
Vibration (non operating)
23.52m/s
2
(10 - 55Hz: constant sweep 1 min X, Y, Z for 1 hour)
Shock
< 196.1 m/s
2
(20G)
Safety Agency Approvals
-
CE, UL60601-1, EN60601-1, IEC60601-1
Safety Class
-
Class I
Altitude
m
3000m
Offload Power Consumption
W
< 0.5W
Conducted & Radiated EMI
EN60601-1-2, FCC Part 18 Class B
Immunity
EN55011
Weight (Typ)
g
460g
Size (WxLxH)
mm
69 x 132 x 39.9
Cable Length
mm
1050mm
AC Input Connector
IEC 320-C14 (Accepts IEC 320-C13)
Output Connector
Kycon KPPx-4P or equivalent, Pins 1 & 2: -Vout, Pins 3 & 4: +Vout
Warranty
yrs
2
Notes:
(1) ErP - Energy Related Products     (2) Californian Energy Commission    (3) EISA - Energy Independence and Security Act of 2007