ACQUISITION - OFFSET, OFST 
Description 
The OFFSET command allows adjustment of the vertical offset of the specified input channel.  
The OFFSET? query returns the DC offset value of the specified channel. 
The maximum ranges depend on the fixed sensitivity setting.  If an out-of-range value is 
entered, the oscilloscope is set to the closest possible value. 
NOTE: The probe attenuation factor is not taken into account in offset adjustments. 
Also, the unit V is optional. 
Command Syntax 
<channel>:OFfSeT <offset> 
<channel>:= {C1, C2, C3, C4} 
<offset>:=
  Refer to datasheet specification for your Teledyne LeCroy oscilloscope at 
www.teledynelecroy.com. 
Query Syntax 
<channel>:OFfSeT? 
Response Format 
<channel>:OFfSeT <offset> 
Example 
The following instruction sets the offset of Channel 2 to -3 V: 
C2:OFST -3V