Analog Inputs
Setting up the inputs
(c) Spectrum GmbH
45
to the input range of ± 1.0 V. After that the four offset settings are set exactely as the offsets to be compensated, but with the the opposite 
sign. The result is, that all four channels match perfectely to the choosen input range.
Overrange bit
With the help of this mode you can additionally record the overrange flag, which is generated by the ADCs. The overrange bit will be stored 
in bit 15 of the samples. If the signal has been out of range this bit will be set to 1, while a 0 indicates that the sample is within the range.
As the overrange bit is generated from sample to sample by the ADC you have to analyze all the recorded 
samples to make sure that the range never has been left.
The sample format corresponding with this mode is explained in the according passage in the chapter relating to the acquisition modes. The 
overrange mode can be enabled by the following register.
Input termination
All inputs of Spectrum’s analog boards can be terminated separately with 50 Ohm by software programming. If you do so, please make sure 
that your signal source is able to deliver the higher output currents. If no termination is used, the inputs have an impedance of 1 Megaohm. 
The following table shows the corresponding register to set the input termination.
 
 
Automatical adjustment of the offset settings
All of the channels are calibrated in factory before the board is shipped. These settings are stored in the on-board EEProm under the default 
settings. If you have asymmetrical signals, you can adjust the offset easily with the corresponding registers of the inputs as shown before.
To start the automatic offset adjustment, simply write the register, mentioned in the following table. Because the adjustment of all the channels 
in all different input ranges can take up some time, it can be useful to adjust only the current input range to safe time.
Before you start an automatic offset adjustment make sure, that no signal is connected to any input. Leave 
all the input connectors open and then start the adjustment. If you adjust all ranges, this can take up some 
time. All the internal settings of the driver are changed, while the automatic offset compensation is in 
progress.
As all settings are temporarily stored in the driver, the automatically adjustment will only affect these values. After exiting your program, all 
calibration information will be lost. To give you a possibility to save your own settings, every Spectrum card has at minimum one set of user 
settings that can be saved within the on-board EEPROM. The default settings of the offset and gain values are read-only and cannot be written 
to the EEProm by the user.
You can easily either save adjustment settings to the EEPROM with SPC_ADJ_SAVE or recall them with SPC_ADJ_LOAD. These two registers 
are shown in the table below. The values for these EEPROM access registers are the sets that can be stored within the EEPROM. The amount 
SpcSetParam (hDrv, SPC_AMP0 ,    1000); // Set up channel0 to the range of ± 1.0 V
SpcSetParam (hDrv, SPC_AMP1 ,    1000); // Set up channel1 to the range of ± 1.0 V
SpcSetParam (hDrv, SPC_AMP2 ,    1000); // Set up channel2 to the range of ± 1.0 V
SpcSetParam (hDrv, SPC_AMP3 ,    1000); // Set up channel3 to the range of ± 1.0 V
SpcSetParam (hDrv, SPC_OFFS0,    -100); // Set the input offset to get the signal symmetrically to 0.0 V
SpcSetParam (hDrv, SPC_OFFS1,     -50);
SpcSetParam (hDrv, SPC_OFFS2,      50);
SpcSetParam (hDrv, SPC_OFFS3,     100);
Register
Value
Direction
Description
SPC_OVERRANGEBIT
201000
read/write
Enables the recording of the ADC overrange bit in data bit 15. If the bit is not zero the signal has 
been out of range.
Register
Value
Direction
Description
SPC_50OHM0
30030
r/w
A „1“ sets the 50 ohm termination for channel0. A „0“ sets the termination to1 MOhm.
SPC_50OHM1
30130
r/w
A „1“ sets the 50 ohm termination for channel1. A „0“ sets the termination to1 MOhm.
SPC_50OHM2
30230
r/w
A „1“ sets the 50 ohm termination for channel2. A „0“ sets the termination to1 MOhm.
SPC_50OHM3
30330
r/w
A „1“ sets the 50 ohm termination for channel3. A „0“ sets the termination to1 MOhm.
SPC_50OHM4
30430
r/w
A „1“ sets the 50 ohm termination for channel4. A „0“ sets the termination to1 MOhm.
SPC_50OHM5
30530
r/w
A „1“ sets the 50 ohm termination for channel5. A „0“ sets the termination to1 MOhm.
SPC_50OHM6
30630
r/w
A „1“ sets the 50 ohm termination for channel6. A „0“ sets the termination to1 MOhm.
SPC_50OHM7
30730
r/w
A „1“ sets the 50 ohm termination for channel7. A „0“ sets the termination to1 MOhm.
Register
Value
Direction
Description
SPC_ADJ_AUTOADJ
50020
w
Performs the automatic offset compensation in the driver either for all input ranges or only the actual.
ADJ_ALL
0
Automatic offset adjustment for all input ranges.
ADJ_CURRENT
1
Automatic offset calibration for the current sampling rate setting.