Spectrum Brands MI.61XX 用户手册

下载
页码 79
42
MI.61xx Manual
Setting up the outputs
Analog Outputs
Setting up the outputs
Output Amplifiers
This arbitrary waveform generator board uses separate output am-
plifiers for each channel. This gives you the possibility to seperately 
set up the channel outputs to best suit your application 
The output amplifiers can easily be set by the corresponding am-
plitude registers.
The table below shows the available registers to set up the output 
amplitude for your type of board.
The amplitude can be changed at any time even if the board is running and outputting a signal to the con-
nectors. The board will not be stopped when changing these settings.
Output offset
In many applications an output of symmetrical sig-
nals is required. But in some cases, depending on 
your application, it can be necessary to generate 
signals that are not symmetrical.
For such cases you can adjust the offset of the out-
puts for each channel seperately.
The figure at the right shows some examples, how 
to set up the offset in combination with different am-
plitudes.
The offset settings can be changed at any time even if the board is running and outputting a signal to the 
connectors. The board will not be stopped when changing these settings.
Register
Value
Direction
Description
Amplitude range
SPC_AMP0
30010
r/w
Defines the amplitude of channel0.
100 mV up to 3000 mV
SPC_AMP1
30110
r/w
Defines the amplitude of channel1.
100 mV up to 3000 mV
SPC_AMP2
30210
r/w
Defines the amplitude of channel2.
100 mV up to 3000 mV
SPC_AMP3
30310
r/w
Defines the amplitude of channel3.
100 mV up to 3000 mV
Register
Value
Direction
Description
Offset range
SPC_OFFS0
30000
r/w
Defines the output’s offset  and therfore shifts the output of channel0.
± 3000 mV in steps of 1 mV 
SPC_OFFS1
30100
r/w
Defines the output’s offset  and therfore shifts the output of channel1.
± 3000 mV in steps of 1 mV 
SPC_OFFS2
30200
r/w
Defines the output’s offset  and therfore shifts the output of channel2.
± 3000 mV in steps of 1 mV 
SPC_OFFS3
30300
r/w
Defines the output’s offset  and therfore shifts the output of channel3.
± 3000 mV in steps of 1 mV