Technical Specifications 
The manufacturer reserves the right to change these specifications without notice.
Symbols which appear on the device
 Attention, consult accompanying documents; 
 Drip proof; 
 Type CF equipment;
 Dangerous voltage; 
 Class II equipment; 
 Start/Stop.
Operating pressure range
4 to 20 cm H
2
O
Dimensions (H x W x D)
4.6" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm)
Weight
3.1 lb (1.4 kg) 
Operating Temperature
+41°F to +95°F (+5°C to +36°C)
Power Supply
Input range for S8 Series II with HumidAire 3i: 100–240V, 50/60Hz; 
110V, 400Hz; 2.5A <140VA (110W) (maximum power consumption). 
Instantaneous peak power consumption <340VA. 
Refer to the DC-12 converter instructions for DC ratings. Only use 
the ResMed DC-12 converter for DC input.
Typical power consumption at 20 cm H
2
O is 39VA or 20.7W.
Operating Humidity
10%–95% non-condensing
Storage and Transport 
Temperature
-5°F to -140°F (-20°C to +60°C)
Storage and Transport 
Humidity
10%–95% non-condensing 
Operating Altitude
Sea level to 8500’ (2591 m)
Electromagnetic
Compatibility
Product complies with all applicable electromagnetic compatibility 
requirements (EMC) according to IEC60601-1-2, for residential, 
commercial and light industry environments. The electromagnetic 
compatibility tables for these ResMed devices can be found on 
www.resmed.com, on the Products page under Service and 
Support. Click on the PDF file for your device.
Air Filter
Two-layered, powder-bonded, polyester non-woven fibre
Air Tubing
Flexible plastic, 6’ 6" (2 m) 
Air Outlet
The 22 mm conical air outlet complies with EN 1281-1
IEC 60601-1 Classifications
Class II (double insulation), Type CF
Supplemental Oxygen
Recommended maximum supplemental oxygen flow: 4 L/min
Housing Construction
Flame retardant engineering thermoplastic
Noise Level
Radiated sound pressure is measured at <24±2 dBA at 10 cm H
2
O and 
3’3.36” (1 m)
 according to ISO17510-1