Your device will ramp to the mask-fit pressure and will remain at this pressure 
for three minutes. 
The following displays on the LCD screen indicating that the mask-fitting feature 
is operating, where: ***** is an excellent fit;****- is a very good fit; ***-- is a 
good fit; while **---, *---- and ----- each require you to adjust your mask.
3
Adjust your mask, mask cushion and headgear until you have a secure and 
comfortable fit. 
After three minutes, the pressure reverts to the set pressure and treatment will 
begin. If you do not wish to wait three minutes, hold down 
 for at least 
three seconds to start treatment immediately. 
If you press 
 for less than three seconds, the unit will return to standby 
mode (the RAMP or SETTLING screen).
Starting Treatment
1
Make sure the power is on. The product name is displayed briefly on the screen, 
then the standby screen displays RAMP (or SETTLING). Press 
 or 
 to 
increase or decrease the ramping (or settling) period. 
2
Fit mask, lie down and arrange the air tubing so that it is free to move if you turn 
in your sleep.
!
CAUTION
Do not leave long lengths of air tubing around the top of your bed. They could 
twist around your head or neck while you are sleeping.
3
To start treatment, press 
 . 
If your clinician has set up the device to start delivering therapy automatically 
using SmartStart™, your device will start when you breathe into your mask.
4
Fit your mask as described in the mask user instructions.
!
WARNING
A mask should not be used unless the S8 Series II device is powered and 
operating properly. 
When treatment begins, if ramp is set, the pressure increases from a low 
pressure to the prescribed treatment pressure over the set period. Ramping is 
designed to help you adjust to an increase in pressure gradually, and enhance 
your comfort. 
 MASK FIT:        *****       
 excelnt