Appendix A: Calibration
15
012-13762D
Step Three: Install the Calibration Rod
The ‘calibration wizard” changes to show an illustration about how to
install the calibration rod.
install the calibration rod.
In addition to the illustration, a Graph display of Position (m) versus
Force (N) opens. A Digits display of Force (N) is part of the Graph dis-
play.
Force (N) opens. A Digits display of Force (N) is part of the Graph dis-
play.
•
Click Next to show Step Four: “Record a smooth data run”.
Step Four: Record a Smooth Data Run
NOTE: To make a calibration that will accurately correct for compli-
ance, it is necessary to calibrate the Materials Testing Machine over the
same range of force and the same conditions you expect to use when
you are testing your samples.
ance, it is necessary to calibrate the Materials Testing Machine over the
same range of force and the same conditions you expect to use when
you are testing your samples.
•
Click Record and collect a smooth run of position versus force
data.
data.
•
When the data collection is finished, click Stop.
REMINDER: If the run of data is not smooth, delete the run and try again.
NOTE: A Polynomial curve fit is automatically applied to the run of position versus
force data.
force data.
•
Click Next to open Step Five, “Polynomial Fit”.
Position (m)
Force (N)
Digits display of
Force (N)
Position versus
force data
Polynomial curve fit