T400 Calibration Procedures 
T400 Ozone Analyzer Operator’s Manual 
192 
 
9.3.2. MANUAL CALIBRATION CHECKS WITH VALVE OPTIONS 
INSTALLED 
Performing the calibration checks on T400 analyzer’s with the Valve option installed is  
similar to that described in Section 9.2, except that the ZERO And  SPAN calibration 
operations are initiated directly and independently with dedicated buttons (CALZ  & 
CALS
). 
 