Theory of Operation 
T400 Ozone Analyzer Operator’s Manual 
282 
 
13.2.3. PARTICULATE FILTER  
The Model T400 Ozone Analyzer comes equipped with a 47 mm diameter Teflon 
particulate filter with a 5-micron pore size.  The filter is accessible through the front 
panel, which folds down to allow access, and should be changed according to the 
suggested maintenance schedule described in Table 11-1. 
 
13.2.4. PNEUMATIC SENSORS 
13.2.4.1. Sample Pressure Sensor 
An absolute value pressure transducer plumbed to the outlet of the sample chamber is 
used to measure sample pressure.  The output of the sensor is used to compensate the 
concentration measurement for changes in air pressure.  This sensor is mounted to a 
printed circuit board next to the internal pump (see Figure 3-5). 
13.2.4.2. Sample Flow Sensor  
A thermal-mass flow sensor is used to measure the sample flow through the analyzer.  
The sensor is located in down stream from the absorption tube but upstream from the 
critical flow orifice.  This sensor is mounted to the same printed circuit board as the 
pressure sensor (see Figure 3-5). 
 
06870C DCN6332