Teledyne API – Model T400 Photometric Ozone Analyzer 
Getting Started 
 
71 
3.5.2. SETTING THE O
GENERATOR LOW-SPAN (MID POINT) OUTPUT 
LEVEL 
To set the ozone LO SPAN (Midpoint) concentration for the IZS O
3
 generator, press:  
SETUP X.X          
0) DAS_HOLD_OFF=15.0 Minutes
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X       
 PRIMARY SETUP MENU
 CFG DAS RNGE PASS CLK
MORE
EXIT
SAMPLE
RANGE=500.0 PPB          O3= XXXX
<TST
TST> CAL
SETUP
SETUP X.X         
SECONDARY SETUP MENU
 COMM
VARS
DIAG
EXIT
SETUP X.X    
ENTER PASSWORD:818
 8
  1
  8
ENTR
EXIT
Continue pressing NEXT until ...
SETUP X.X          
4) O3_GEN_LOW1=100.0 PPB
PREV
NEXT
JUMP
EDIT
PRNT EXIT
Sets LOW SPAN Point for 
RANGE1.
To Set the LOW SPAN point for 
RANGE2 in DUAL or AUTO
range modes …
Press NEXT button once more to 
select O3_GEN_LOW2
 then continue as shown.
SETUP X.X          
4) O3_GEN_LOW1=100.0 PPB
0
1
0
0
.0
ENTR EXIT
EXIT discards the new 
setting
ENTR accepts the 
new setting
Toggle buttonss to change setting
Only Values from  0 to 1500 will be accepted..
A value of 0 turns the lamp OFF.
The ENTR key will disappear if an invalid setting 
is attempted.
 
06870C DCN6332