Teledyne API - T100 UV Fluorescence SO2 Analyzer
SETUP Menu
99
The analyzer multiplies the measured gas concentrations with this dilution factor and
displays the result.
displays the result.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Once the above settings have been entered, the instrument needs to be
recalibrated using one of the methods discussed in Section 9.
recalibrated using one of the methods discussed in Section 9.
5.5. SETUP – PASS: PASSWORD PROTECTION
The menu system provides password protection of the calibration and setup functions to
prevent unauthorized adjustments. When the passwords have been enabled in the PASS
menu item, the system will prompt the user for a password anytime a password-
protected function (e.g., SETUP) is selected. This allows normal operation of the
instrument, but requires the password (101) to access to the menus under SETUP. When
PASSWORD is disabled (SETUP>OFF), any operator can enter the Primary Setup
(SETUP) and Secondary Setup (SETUP>MORE) menus. Whether PASSWORD is
enabled or disabled, a password (default 818) is required to enter the VARS or DIAG
menus in the SETUP>MORE menu.
prevent unauthorized adjustments. When the passwords have been enabled in the PASS
menu item, the system will prompt the user for a password anytime a password-
protected function (e.g., SETUP) is selected. This allows normal operation of the
instrument, but requires the password (101) to access to the menus under SETUP. When
PASSWORD is disabled (SETUP>OFF), any operator can enter the Primary Setup
(SETUP) and Secondary Setup (SETUP>MORE) menus. Whether PASSWORD is
enabled or disabled, a password (default 818) is required to enter the VARS or DIAG
menus in the SETUP>MORE menu.
Table 5-1: Password Levels
PASSWORD
LEVEL
MENU ACCESS ALLOWED
Null (000)
Operation
All functions of the main menu (top level, or Primary, menu)
101
Configuration/Maintenance
Access to Primary and Secondary SETUP Menus when PASSWORD is
enabled
enabled
818
Configuration/Maintenance Access to Secondary SETUP Submenus VARS and DIAG whether
PASSWORD is enabled or disabled.