Quest Diagnostics QUESTEMP 36 用户手册

下载
页码 14
 
21
Remote Sensor Bars:  
2 x 15pin D-sub jacks are located on the side of the unit for plugging in 1 or 
2 additional sensor bars by using remote cables up to 200 feet (61m).  The 
top sensor bar can also be remote with a cable.   
 
Power Options:  
9V alkaline,  7.2V NiMH rechargeable pack (charged in the unit), or AC 
adapter wall power cube (AC adaptor will operate the unit or recharge the 
NiMH battery pack) 
 
Battery Life:  
9V alkaline: 140 hours 
Rechargeable Nickel Metal Hydride: 300 hours 
(Adding additional sensor bars to the QUESTemp
°36 reduces battery life.) 
 
Charge Time (NiMH Battery Pack):  
16 hours (charge in the unit) 
 
 
Safety Approvals: 
ETL, cETL: Class I,II,III Groups A,B,C,D,E,F,G, Temperature code T3  
DEMKO: EEX ia IIC, T3 
CE mark  
 
AIR PROBE ACCESSORY 
 
Range: 
 
 0 - 20 meters per second.  0.1m/s increments 
 
Sensor: 
 
Omni directional heated thermistor 
 
Accuracy: 
 
+/- (0.1 m/s + 4%) of measurement value 
 
Battery Life: 
 
6 - 8 hours for fully charged NiMH battery 
 
Charge Time: 
 9 
hours 
 
22 
 
16.  APPENDIX A : HEAT EXPOSURE TABLES 
 
ACGIH 
Screening Criteria for Heat Stress Exposure.   WBGT values in 
°C (°F). 
 
 
 
 
 
 
 
Acclimatized 
Work 
 
 
 
 
 
 
 
 
 
 
 
Very 
Demands  Light   Moderate Heavy   Heavy 
 
100% 
Work 
 
 29.5 
  27.5 
  26.0 
   
    (85.1)  (81.5)  (78.8) 
 
75% Work;     
30.5  
 
28.5  
 
27.5 
25% 
Rest 
 
 
 (86.9)  (83.3)  (81.5) 
 
50% 
Work; 
 
 31.5 
  29.5 
  28.5 
  27.5 
50% Rest        
(88.7)   
(85.1)   
(83.3)   
(81.5) 
 
25% 
Work; 
 
 32.5 
  31.0 
  30.0 
  29.5 
75% Rest        
(90.5)   
(87.8)   
(86.0)   
(85.1) 
 
 
 
 
 
 
 
 
Unacclimatized 
Work 
 
 
 
 
 
 
 
 
 
 
 
Very 
Demands  Light   Moderate Heavy   Heavy 
 
100% 
Work 
 
 27.5 
  25.0 
  22.5 
   
    (81.5)  (77.0)  (72.5) 
 
75% 
Work; 
 
 29.0 
  26.5 
  24.5 
25% Rest        
(84.2)   
(79.7)   
(76.1) 
 
50% 
Work; 
 
 
 30.0 
  28.0 
  26.5 
  25.0 
50% 
Rest 
 
 
 
 (86.0)  (82.4)  (79.7)  (77.0) 
 
25% 
Work; 
 
 
 31.0 
  29.0 
  28.0 
  26.5 
75% 
Rest 
 
 
 
 (87.8)  (84.2)  (82.4)  (79.7) 
 
from "American Conference of Governmental Industrial Hygienists - Threshold 
Limit Values and Biological Exposure Indices for 2001" 
Reprinted with permission from ACGIH