equipment, circuits powered by regulated low-voltage sources, and 
electronics.
Installation Category II is for measurements performed on circuits 
directly connected to the electrical distribution system. This category 
refers to local-level electrical distribution, such as that provided by a 
standard wall outlet (e.g., 115 V for U.S. or 230 V for Europe). 
Examples of Installation Category II are measurements performed on 
household appliances, portable tools, and similar products.
Installation Category III is for measurements performed in the building 
installation at the distribution level. This category refers to 
measurements on hard-wired equipment such as equipment in fixed 
installations, distribution boards, and circuit breakers. Other examples 
are wiring, including cables, bus-bars, junction boxes, switches, 
socket-outlets in the fixed installation, and stationary motors with 
permanent connections to fixed installations.
Installation Category IV is for measurements performed at the primary 
electrical supply installation (<1,000 V). Examples include electricity 
meters and measurements on primary overcurrent protection devices 
and on ripple control units.
Connection Overview
Figure 2-1 illustrates the sequence for connecting and getting started with 
the NI 1450.