Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 42 
 
www.PressureSystems.com 
Internal firmware limits calculated gains to values are software limited to values between 0.0 
and 100.0.  Any calculated value outside of this range will result in the gain coefficient being set 
to 1.00. 
 
Note
 
The calculated gain values from the latest ‘Z’ command will be lost 
when the module is powered off.  To  save these gain
 terms to each 
transducer’s non-volatile memory, refer to the Set Operating Options
 
(‘w’) command (index 09).
 
 
Example: 
 
Send TCP/IP command to a Model 9116 module (via its open socket) to  
 
calculate and set gain coefficients for channels 8 through 4. Instruct the module  
 
to use 14.8890 psi as the applied pressure instead of each transducer’s full-scale  
 value: 
 
 
 
Z00F8 14.8890” 
 
Read response, containing the new gain values (also stored in the module’s  
volatile main memory): 
 
 
1.000212 1.000269 1.000437 1.000145 .999670” 
 
Actual gain values are returned in the above response as decimal ASCII strings,  
each preceded by a space character. From left-to-right: they are for channels 8,  
7, 6, 5, and 4.