Table 2 Analog Outputs
Parameter
Name
Oxygen
Default
Possible
Options
Options
Possible
Ranges
OUTPUT 1
MODE
MODE
O2
0–20.9%
4-20mA
4-20mA
O2, CARBON,
DEWPT, TEMP, LIN,
PROG
DEWPT, TEMP, LIN,
PROG
O2 = 0 – 9999
%C = 0.00 – 2.55
DP = -99.9 – 212.0
Temp = -999 – 3000
LIN = -999 – 9999
PROG = 0 – 4095
%C = 0.00 – 2.55
DP = -99.9 – 212.0
Temp = -999 – 3000
LIN = -999 – 9999
PROG = 0 – 4095
OUTPUT 2
MODE
MODE
TEMP
800-3000°F
4-20mA
4-20mA
O2, CARBON,
DEWPT, TEMP, LIN,
PROG
DEWPT, TEMP, LIN,
PROG
O2 = 0 – 9999
%C = 0.00 – 2.55
DP = -99.9 – 212.0
Temp = -999 – 3000
LIN = -999 – 9999
PROG = 0 – 4095
%C = 0.00 – 2.55
DP = -99.9 – 212.0
Temp = -999 – 3000
LIN = -999 – 9999
PROG = 0 – 4095
NOTE: SEE PAGE 4 FOR TYPICAL CALIBRATION VALUES.
Calibration
The Smart Transmitter is factory calibrated. The calibration can be verified once a year or
according to customer calibration schedules. The instrument should be returned to the
factory if calibration is required.
according to customer calibration schedules. The instrument should be returned to the
factory if calibration is required.