Figure 1-1 shows the relation between frequency of floor vibration and vibration transmittance of 
the instrument.    Since the vibration transmittance is maximum at the resonance point (about   
2 Hz) of the anti-vibration mount provided with the instrument, the allowable amplitude is 
minimum at this point.   
 
NOTICE:  
Please consult with us if the floor vibration exceeds the allowable values. (Vibration 
should be measured in advance if it is expected to cause a disturbance.)   
 
Low frequency
vibration range
Frequency : fn
5 Hz
Resonance point
T
rans
mittanc
e T