2.3.7 
2 - 47 
2.3.7.11  DATA ENTRY Window 
 
Use for drawing texts and graphics on the scanning image. 
 
 
 
Fig. 2.3-35    DATA ENTRY Window 
 
Refer to <3.9.8    Data Entry Function>. 
 
 
2.3.7.12  TILT COMPENSATION Window 
 
Specimen tilting causes defocusing and magnification change in the tilt direction.   
Tilt compensation is the tool to correct the two phenomena. 
 
 
 
 
Fig. 2.3-36    TILT COMPENSATION Window 
 
(1) Dynamic focus 
Checking the Dyn.Focus check box enables dynamic focusing (the function that varies the 
focusing in tandem with the scanning process).    The adjustment slider and the micro-
adjustment button can be used to make adjustments so that the image will be in focus over 
the entire screen. 
The possible range of adjustment is ±100(±70°) so that compensation can be made in both 
positive and negative directions of tilt.    These numbers are index values, not angles. 
Adjustments should be made by viewing the focus on the image. 
 
Dynamic focus check box
 
Dynamic focus display/entry box 
Dynamic focus micro adjustment button
 
Dynamic focus adjustment slider 
Tilt-magnification compensation check box 
Tilt-magnification compensation display/entry box
 
Tilt-magnification compensation micro adjustment button
 
Tilt-magnification compensation adjust slider