3.8.3 
3 - 113 
3.8.3  Withdrawing the Specimen 
 
The following procedures can be employed to withdraw a specimen from the specimen chamber: 
 
(1)  On the EVAC panel located on the front side of the main unit, press the AIR button.   
Alternatively, click the AIR button located in the upper right section of the Control Panel.   
This introduces air into the specimen chamber.    Wait approximately 100 seconds until the 
specimen chamber is completely at the atmospheric pressure. 
 
 
 
 
 
 
EVAC panel 
Control Panel 
 
Fig. 3.8-2    AIR Button (letting air into the specimen chamber) 
 
(2)  Grasp the handle on the front side of the specimen stage.    Pull the specimen stage toward 
you, and take out the specimen.    When removing the specimen, be sure to put on clean 
gloves; do not use a bare hand. 
 
 
 
 
 
 
Fig. 3.8-3    Pulling Out the Stage 
Fig. 3.8-4    Removing the Specimen 
 
IMPORTANT:  Do not touch the operation parts such as gears and driving screws and note that 
clothes are never placed when you bring the hand close to the movement 
mechanism in the specimen stage by the specimen exchange etc.