  For Safe Handling 
SAFETY - 20
 
Precautions on Proper Operation (cont’d)
 
 
Transportation and Relocation of Instrument 
 
(a) The transport of the system, which carries considerable risk, 
should never be attempted by a customer.    Any system transport 
should be performed with the assistance and supervision of 
Hitachi-approved engineers. 
(b) Do not lift the instrument by holding the table.    The strength of 
table fitting is not sufficient for bearing the weight of display unit, 
approximately 120 kg.    Should the table be lifted, the display unit 
might slip off and crash.    Hence, it is recommended to remove 
the table and transport the display unit independently when 
moving the instrument. 
 
Ambient Temperature and Humidity 
 
Even when the system is not running, the room in which the system 
is installed should be maintained under the following environmental 
conditions: 
Room temperature:  15
℃ to 30℃ 
Humidity:  70
%  or lower (no condensation) 
If the system is not used for a long time, the specimen chamber 
should be evacuated instead of being exposed to the atmosphere.