4.4  Troubleshooting 
4 - 26 
4.4  Troubleshooting 
 
4.4.1    The Evacuation System does not Run 
 
If the evacuation system fails to run or the evacuation system does not proceed, the following 
items should be checked: 
 
(1) 
Is there an error?   
If both the EVAC/AIR switches on the EVAC Panel are blinking and the buzzer is sounding, 
it is an indication that an error has occurred.    If this happens, check the error message on 
the PC operation screen. If an error message is displayed, the corrective action described 
in the list of error messages in the Appendix should be taken. 
 
(2) Is 
the 
PUMP light on the EVAC Panel either lit or blinking?   
In the steady-state operation condition, the pump is running normally if the PUMP light is on. 
During system startup, the PUMP light begins to blink after the compressor air pressure 
has reached a prescribed level and values are set, and the pace of blinking picks up as the 
rpm for the turbo molecular pump increases.    When the rpm for the turbo molecular pump 
has reached a prescribed level and the pump goes into the normal running condition, the 
PUMP light becomes solidly lit.   
Normally, the turbo molecular pump reaches a set rpm level and the PUMP light becomes 
solidly lit within 10 minutes after the power is turned on.    If the PUMP light fails to blink or 
the blinking condition lasts more than 10 minutes, and if an error message is displayed on 
the PC operation screen, the corrective action described in the list of error messages in the 
Appendix should be taken.    If an error message does not come up, call the service 
engineer.