Hitachi Microscope & Magnifier S-3400N 用户手册

下载
页码 466
6.3.2 
6 - 17 
6.3.2.2  Acceleration Voltage 
 
(1)  Monitoring/Setting of Acceleration voltage 
 
 Property 
 
 
HV ON/OFF monitoring 
 
 
HVOn 
as Long 
Read Only 
0: HV–OFF 
   
 
1: 
HV-ON 
 
 
 
Acceleration voltage monitoring 
 
 
Vacc 
as Long 
Read Only 
0: HV–OFF 
 
 
 
 
300 to 30000 :    0.3 – 30.0 (kV) 
 
 Method 
 
 
HV ON/OFF setting 
 
 
SetHVOn (HVON as Long) as Long 
 
 
 
HVON 
0: HV-OFF  1: HV-ON 
 
 
 
Return 
0: Normal receiving 
 
 
 
 
Excluding 0: Abnormal receiving 
 
 
 
Comment: 
The method is finished at completion of 
command analysis. 
The completion of acceleration voltage 
setting is notified by the HVOnFinished event. 
If the return value is not 0, the HVOnFinished 
event is not issued. 
Please do not use other method until 
receiving the HVOnFinished event.