Hach 900 MAX 用户手册

下载
页码 160
Section 5
Page 75
8990options.fm
Dissolved Oxygen Probe
5.4.3 Dissolved Oxygen Probe Temperature Programming
1.
From the Main Menu, select 
OPTIONS>ADVANCED OPTIONS>DATALOG
2.
Highlight Select Inputs using the 
UP
 and 
DOWN 
keys and press 
SELECT
.
3.
Highlight D.O. Temp. using the 
UP
 and 
DOWN 
keys, then press 
SELECT
.
4.
Cycle between Logged and Not Logged, then press 
ACCEPT
.
5.
Enter a logging interval using the numeric keypad, then press 
ACCEPT
Valid logging intervals are shown on the status bar.
6.
Select temperature units (°C, °F) using 
CHANGE CHOICE
. Press 
ACCEPT
.
5.4.4 Dissolved Oxygen Probe Calibration
The Dissolved Oxygen Probe is shipped without an electrolyte or membrane 
installed. Install these items before initial use.
1.
Connect a suitable power supply to the sampler, and power up the unit by 
pressing the 
ON
 button.
2.
From the Main Menu, select 
OPTIONS > ADVANCED OPTIONS > 
CALIBRATION > DO.
3.
Enter the ambient air temperature (the current reading is displayed for 
reference) using the numeric keypad.
4.
Enter the elevation above sea level for the specific location.
5.
Enter the membrane thickness. The operation of the sampler will be 
affected by membrane thickness for the oxygen sensor.
Note: The membrane of a charged 
sensor must be kept moist. If the 
membrane is allowed to dry 
completely, the electrolyte film 
between the membrane and the 
platinum will evaporate, 
destabilizing the sensor. If the 
sensor will be out of water for more 
than 30 minutes, put a small amount 
of water in the silicon soaking cap, 
and install it over the protective 
guard. Lift the edge of the cap to 
break the seal as it is being 
removed. This will prevent a vacuum 
from forming inside the soaking cap 
while it is being removed which can 
result in the membrane becoming 
stretched.
For general purpose applications, the 1-Mil membrane is standard. This 
membrane allows measurements in the 0 to 20 ppm range of dissolved 
oxygen and provides the best response time and durability.
The 2-Mil membrane can be used to measure up to 40 ppm dissolved oxygen. 
Its increased thickness slows the response time of the sensor, but this 
membrane has increased resistance to cuts and tears. For this reason, it is 
recommended for use in wastewater aeration basins where solids in the water 
are in rapid motion.
6.
Enter the chlorinity (salinity) of the flow stream (typical wastewater is zero, 
sea water is higher).
7.
Place the D.O. probe in open air and press any key. The sampler will wait 
for the reading to stabilize before storing the calibration value. The screen 
will automatically return to the calibration menu.
Calibrating the D.O. Temperature
1.
Place the probe and the thermometer in a liquid.
2.
Wait for the temperature reading to stabilize.
3.
Enter the actual temperature of the liquid.