Section 6
Page 89
8990cm.fm
Alarm Relays
7.
Press any key to set the output to 20.00 mA dc.
8.
Measure the current on the selected output using the multimeter and 
enter the measured value using the numeric keypad. Press 
ACCEPT
 to 
complete the calibration.
By entering the measured current values, the microprocessor will 
electronically adjust the outputs to compensate for the difference between the 
measured values and the expected values.
Figure 24
Calibration with the Meter in the Loop
Figure 25
Calibration with the 4–20 mA Device Disconnected from the Loop
6.4 Alarm Relays 
Note: Current to the relay contacts 
must be limited to 5 amps. A means 
to remove power from the relays 
locally in case of an emergency or 
for servicing the product must be 
provided by the user. This can be 
accomplished with an external 
switch and a 5-amp fuse or with a 
switched 5-amp circuit breaker. 
Four alarm relay outputs are available as a factory installed option. The relays 
are mounted in an external NEMA 4X enclosure for installation to a wall or 
panel. Alarm contacts are rated for 10 amps at 240 V ac (resistive load). 
The alarm wiring can be sized according to the load being used. The relay 
connector will accept wire sizes from 18–12 AWG with a rating of 300 V, 80 °C 
minimum. Do not use wire smaller than 18 AWG.
For relay pin assignments refer to 
 and 
Chart
Recorder
Multimeter
900 MAX
Current Loop
Multimeter
900 MAX