Maintenance
7.2 Lamp replacement
CAUTION
The UV lamp generates UV light. Do not look directly at an 
operating lamp without wearing UV protective eye glasses.
The Lamp compartment is on the left side behind the display and is 
provided with ventilation on the back side. The tungsten and 
deuterium (UV) lamp are installed in the lamp compartment. On the 
back side a fan is installed for cooling of electric components.
The ventilation system operates automatically.
Refer to 
 for instructions on replacing the tungsten 
lamp and deuterium lamp (UV).
7.2.1 Tungsten lamp replacement
1. Switch the instrument off.
2. Unplug the power cord.
DANGER
The instrument must remain switched off until the change of 
lamp has been completed.
WARNING
Burn Hazard. Wait until the lamp cools down. Contact with the 
hot lamp can cause burns.
3. Use a screwdriver to remove the cover from the back of the 
instrument (the screws may be slotted or cross-headed).
4. Place the cover and the attached fan carefully beside the 
instrument (take special care with the fan cable).
5. Push the spring (
, item 
) up and remove 
the tungsten lamp (
, item 
) from the lamp 
compartment.
6. Unplug plug contact (
) from the tungsten lamp.
Important Note: The lamp should only be held by the lamp socket. 
Avoid touching the glass. 
7. Push plug contact firmly onto the new tungsten lamp. 
8. Insert the tungsten lamp into the lamp compartment. Push the 
spring down. 
9. Check that the spring and the lamp socket are positioned 
correctly. 
10. Use a screwdriver to screw the back cover onto the instrument 
again.
11. Plug in power supply again.