Section 3
Installation
WARNING
Electrical and Fire Hazards. Use only the provided power 
supply. Only qualified personnel should conduct the tasks 
described in this section of the manual.
3.1 Unpack the instrument
The DR 5000 Spectrophotometer comes packaged with the 
following items:
DR 5000 Spectrophotometer
Mains power cable
Multi Cell Holder
DR 5000 User Manual
CD-ROM containing the HACH and HACH LANGE procedures
Note: If any of these items are missing or damaged, contact the 
manufacturer or a sales representative immediatly.
3.2 Operating environment
The following conditions are necessary to ensure correct 
instrument operation and accurate results: 
Place the instrument firmly on an even surface. Do not push 
any objects under the instrument.
Maintain an ambient temperature of 10 to 40 ºC (50 to 104 ºF) 
for proper instrument operation.
The relative humidity should be less than 80%; moisture should 
not condense on the instrument.
Leave at least a 15 cm (6 in.) clearance at the top and on all 
sides for air circulation to avoid overheating of electrical parts.
Do not operate or store the instrument in extremely dusty, 
damp or wet locations.
Keep the surface of the instrument, the cell compartment and 
all accessories clean and dry at all times. Splashes or spills on 
and in the instrument should be cleaned up immediately (see 
).
Important Note: Protect the instrument from temperature 
extremes, including heaters, direct sunlight and other heat sources.