Hach DR 5000 用户手册

下载
页码 136
47
Standard Operations
5. Press Adjust to enable the Standard Adjust. The Standard 
Adjust icon will appear.
Note: The adjustment must be within certain limits, which vary with 
each program. The allowable percentage is shown after "Adjustment“.
Note: When a Standard Adjust is in effect, the Standard Adjust icon 
will appear on the display.
5.4.7 Set the chemical form
Some Stored Programs allow a variety of chemical forms to
be selected.
Press the unit (e.g. mg/L) or the chemical representation of the 
evaluation form (e.g. Al
3+
). A list of available evaluation forms is 
displayed. Select the required form by pressing the corresponding 
entry in the list.
Note: When the program is exited, the evaluation form reverts to the 
standard setting.
An alternative way of changing the standard setting:
1. Press Options>More>Chemical Form.
2. Select the Chemical Form.
Note: The stoichiometric conversion of the measurement result is 
carried out automatically.
Note: The selected Chemical Form will appear on the display. Test 
results will be calculated and displayed in this chemical form.
5.4.7.1 Change of the default setting of the chemical form
1. Insert the sample cuvette or blank (depending on the working 
procedure) into the cell compartment.
2. In the result display, press Options>More>Chemical Form.
3. A list of available evaluation forms appears. Select the new 
default setting.
4. Press Save as Default. 
The current result and all further measurements will be displayed in 