Advanced Operations
e. Does the test require the sample to be at a certain 
temperature?
f.
Was the pH of the sample in the correct range?
g. Is the pipette volume correct?
2. Check the used reagents by repeating the standard addition 
procedure with freshly prepared reagents.
If none of these measures resolves the problem, please contact the 
manufacturer or a sales representative.
6.3.1 Complete a standard addition
Carry out the standard addition in accordance with the 
corresponding procedure.
There are two different methods:
Peak volume (Standard addition):
Defined volumes of a standard solution are added step by step to 
an already analyzed sample. The sample is measured again after 
each addition.
Sample volume:
A defined volume of a standard solution of known concentration is 
added to the volume of sample specified in the procedure and the 
sample is measured after each addition. In most cases three 
different standard solutions are prepared and the procedure is 
repeated for each of them.
Note: The units and chemical forms used for the sample are used for the 
standard solutions. Be careful to ensure that to use the correct units for 
subsequent entries.
Detection rate
Conclusion
100%
The originally used reagents were defective. 
Check the standard solution: 
Repeat the standard addition procedure with a 
freshly prepared standard solution.
Detection rate
Conclusion
100%
The originally used 
standard solution was 
defective.