Hach DR 5000 用户手册

下载
页码 136
89
Advanced Operations
Reading mode:
1. To highlight the required mode, start by pressing Reading 
mode.
2. Select the required mode, then press OK, then Return to return 
to the result display.
6.4.2 Take single wavelength measurements (single reading)
1. Insert the blank cuvette/cell into the cuvette/sample cell holder. 
Press Zero.
Note: The Read key is only active after the zero measurement has 
been completed.
2. Insert the sample cuvette/cell into the cuvette/sample cell 
holder. Press Read.
3. For data storage, se
6.4.3 Take single wavelength measurements (continuous readings)
1. Insert the blank cuvette/cell into the cuvette/sample cell holder. 
Press Zero.
Note: In the reading mode ”Continuous“ only the Zero key is shown to 
start the reading. The reading sequence is started automatically.
2. Insert the sample cuvette/cell into the cuvette/sample cell 
holder.
3. Press Options and then the Store icon to store the displayed 
data in the Data Log.
Note: For data storage, se
.