This product is a variable height, adjustable mattress
platform, which will provide comfort and convenience
for residents/patients and caregivers in long term
care settings.
The MAXIMUM SAFE WORKING LOAD for the
APS 9000 Bed with weight evenly distributed, including
bedding, resident/patient, support surface, and all
accessories, is 600 lbs. 
NEVER operate the bed if a Power Cord or Plug is
damaged or not working properly. Contact qualified
Service Personnel for examination and repair. Always
unplug the Power Cord when performing any mainte-
nance on the bed.
DO NOT open assemblies such as the Actuators, Hand
Control, or Control Box. If unauthorized personnel
perform work on these components, the manufacturer’s
warranty becomes void.
DO NOT use unauthorized parts, accessories, or
adaptors other than those specified/authorized by
GF Health Products, Inc.
When operating the High/Low, Knee, or Back Functions
of the bed, ALWAYS ensure that the confined individual
is positioned properly within the confines of the bed.
DO NOT let any extremities protrude over the side or
between the bed rails when performing these functions.
DO NOT lower the bed when objects are beneath it.
Failure to inspect under the bed can result in damage
to property or personal injury.
The bed’s Pendant Cord MUST BE ROUTED AND
SECURED PROPERLY to ensure it does not become
entangled and eventually severed during use. Also
make sure all electrical cords DO NOT get tangled
around the bed, side rails, or legs during transport or
normal operation of the bed.
When using nasal- or masked-type administering
equipment, all oxygen or air tubing MUST BE ROUTED
AND SECURED PROPERLY to ensure that the tubing
does not become entangled and eventually severed
during the normal operation of the bed.
The bed should ALWAYS be left in its lowest position
when unattended to reduce the risk of injury while getting
in or out of the bed.
Keep all moving parts free of obstructions (i.e. blankets/
sheets, heating blankets/pads, wiring, etc.)
DO NOT use the assist devices as push handles for moving
the bed. Assist devices can be deformed or broken if
excessive side pressure is exerted. Assist devices are not
meant for patients considered as high risks for entrapment
(i.e. patients with pre-existing conditions such as 
confusion, restlessness, lack of muscle control, altered
mental status, either organic or medicinal, or a combi-
nation thereof). Additional safety measures should be
considered for such high-risk patients. 
NEVER permit more than one (1) person on/in the bed
at any time.
Body weight should be evenly distributed over the
sleeping surface of the bed. DO NOT allow the patient
to lay, sit, or lean in such a way that their entire body
weight is placed only on the raised head or foot sections
of the bed. This especially applies when repositioning
or transferring a patient in or out of the bed. Increased
risk may occur when the patient’s size and/or weight are
inappropriate for the bed’s dimensions or weight capacity.
The bed is intended for use, storage, and transport within
a temperature range of -40˚C to +60˚C. It has a water-
resistance rating of IPX4 and IS NOT to be power-
washed or submersed. 
!
!
!
!
!
!
!
!
!
!
!
!
!
!
!
operating Conditions
Operation of the bed is based on the following conditions:
Ambient Temperature of +10˚C to +40˚C; Relative Humidity
of 30% to 75% (Non-condensing); Atmospheric Pressure of
700hPa to 1060hPa; and a Splash Protection of IEC 60529.
storage
Storage of the bed is based on the following conditions:
Ambient Temperature of -10˚C to 70˚C; Relative Humidity
of 10% to 100%; and an Atmospheric Pressure of 500hPa
to 1060hPa.
Radio Frequency Interference (RFI)
RFI influences most electronic equipment. Caution should be
exercised with regard to the use of portable communications
equipment in the area around such equipment. If RFI causes
erratic behavior, shut the bed off immediately. Leave it off
while the transmission is in progress. 