Gossen Metrawatt 751 用户手册

下载
页码 48
22
GOSSEN METRAWATT GMBH
9.6
Alternating / Direct Voltage U
AC/DC
 – Max. 253 V (feature F01)
Direct, alternating and pulsating voltages of up to 253 V can be measured 
between the N (2) and L (3) connector jacks.
Select the 
U
AC/DC
 
measurement and start.
9.7
Resistance R (feature F01)
Resistance of up to 150 k
Ω can be measured between the PE (1) and N 
(2) connector jacks.
Select the 
measurement and start.
9.8
Measurements with Accessories (feature F01)
9.8.1
Alternating Current I
Z
 with Clip-On Transformer (feature F01)
Connection 
Alternating current can be measured in two measuring ranges 
(1 mA 
… 10 A ~, 1 A … 100 A~) with a clip-on current-voltage 
transformer connected to the N (2) and L (3) jacks (e.g. the WZ12C).
Start the 
I
measurement.
Select measuring range.