FB300-2  MANUAL 
Page 2 
7608001 Rev A   8/02 
Safety Considerations 
Read and understand this manual completely before attempting to set up or 
use this instrument. 
This equipment has been designed and tested to conform to IEC1010-1 safety standards, as 
applicable to laboratory instrumentation.  This applies only to the FB300 when used as specified 
in the documentation, in its intended applications, with Fisher Biotech approved electrophoresis 
apparatus only.  Usage in any other manor may not provide similar performance or safety 
protection. 
This equipment is provided with a 3-conductor, grounded AC line cord.  The protective earth 
ground is necessary for safe operation.  Do not use any other AC line cord with this instrument. 
The FB300 is a high voltage power supply capable of generating dangerous levels of voltage and 
current during operation.  Exercise caution when working around and with the electrical 
connections of this equipment.  Always check electrical connectors, wires, and associated 
apparatus for any signs of wear or damage before using with this equipment.  Be sure to use only 
electrophoresis equipment that is suitably rated for the voltage and current capabilities of the 
FB300 power supply. 
The output of the FB300 power supply is intended for connection to electrically isolated 
electrophoresis apparatus only.  Use only with electrically isolated electrophoresis apparatus with 
minimum isolation of 600V.  Do not connect any terminal of the FB300 output to earth ground.  
This may impair the safety protection provided by the equipment, or cause equipment damage. 
The high voltage output of the FB300 power supply takes some amount of time to decay when 
unloaded or lightly loaded.  Wait a minimum of 60 seconds after stopping a run before touching 
the power supply leads. 
This equipment has a protective ground leakage current less than 0.5 mA using test methods 
defined in CSA1010 and CSA151. 
This equipment is for indoor use only.