FB1000-2 MANUAL 
Page 2 
7608177 Rev. A  8/9/02 
Safety Considerations 
Read and understand this manual completely before attempting to set up or 
use this instrument. 
This equipment has been designed and tested to conform to IEC1010-1 safety standards, as 
applicable to laboratory instrumentation.  This applies only to the FB1000 when used as specified 
in the documentation, in its intended applications, with Fisher Biotech approved electrophoresis 
apparatus only.  Usage in any other manor may not provide similar performance or safety 
protection. 
This equipment is provided with a 3-conductor, grounded AC line cord.  The protective earth 
ground is necessary for safe operation.  Do not use any other AC line cord with this instrument. 
The FB1000 is a high voltage power supply capable of generating dangerous levels of voltage and 
current during operation.  Exercise caution when working around and with the electrical 
connections of this equipment.  Always check electrical connectors, wires, and associated 
apparatus for any signs of wear or damage before using with this equipment.  Be sure to use only 
electrophoresis equipment that is suitably rated for the voltage and current capabilities of the 
FB1000 power supply. 
The output of the FB1000 power supply is intended for connection to electrically isolated 
electrophoresis apparatus only.  Use only with electrically isolated electrophoresis apparatus with 
minimum isolation of 1000V.  Do not connect any terminal of the FB1000 output to earth ground.  
This may impair the safety protection provided by the equipment, or cause equipment damage. 
The high voltage output of the FB1000 power supply takes some amount of time to decay when 
unloaded or lightly loaded.  Wait a minimum of 60 seconds after stopping a run before touching 
the power supply leads. 
This equipment has a protective ground leakage current of 0.5mA at 120VAC, and 1mA at 
230VAC.  While significantly below the 3.5mA limitation for laboratory instrumentation, some 
applications and locations require a leakage current below 0.5mA.  Check the specific 
requirements of your application before using this equipment. 
This equipment is for indoor use only.