Agilent Technologies Microwave Oven 53150A 用户手册

下载
页码 142
Appendix B  Messages
Error Messages
B-4
Operating Guide
B
Error Messages
Table 3-3 lists and describes messages that are generated by the Counter 
during Self-Test or during operation to indicate that a problem has been 
detected. These messages are displayed on the Counter’s front-panel 
display and are also sent via the RS-232 serial output (note that, in many 
cases, the exact message text that is displayed on the front panel is a 
condensed form of the message that is sent via RS-232).
Table 3-3. Error Messages
Message
Display RS-232
Description
12V FAIL
X
X
The +12 VDC output from the power 
supply is not within specifications.
–12V FAIL
X
X
The –12 VDC output from the power 
supply is not within specifications.
–5V FAIL
X
X
The –5 VDC output from the power 
supply is not within specifications.
ADC FAIL
X
X
A failure was detected in the ADC.
B1 SIGNAL PATH FAIL
X
A failure was detected in the Channel 1 
signal path.
PATH FAIL
X
B1 THRESHOLD FAIL
X
A failure was detected in the Channel 1 
threshold circuit.
THRS FAIL
X
B2 RF THRESHOLD FAIL
X
A failure was detected in the Channel 2 
RF threshold circuit.
THRS FAIL
X
B2 THROUGH-PATH 
THRESHOLD FAIL
X
A failure was detected in the Channel 2 
through-path threshold circuit.
THRS THRU
X
B2 HETERODYNE PATH 
THRESHOLD FAIL
X
A failure was detected in the Channel 2 
heterodyne-path threshold circuit.
THRS HET
X
EEPROM FAIL - WRITE
X
A failure was detected while writing to 
EEPROM.
EEP WRT FAIL
X