Chapter 1  Getting Started
Operating the Counter
1-22
Operating Guide
1
Measuring Power (Channel 2 Only)
The Agilent 53150A/151A/152A Counter can measure signal power (in the 
same frequency ranges as for frequency measurements) on Channel 2. 
The example procedure for measuring power in the following diagram 
assumes that the Counter has previously been set up to measure a 25 GHz 
input on Channel 2.
NOTE
When the power of a signal applied to the Channel 2 input exceeds the 
maximum rated power for the instrument, the Counter displays HI.
CAUTION
The Channel 2 input path circuits contain sensitive GaAs semiconductors. 
To prevent damage to these components, always adhere to standard ESD 
(Electro-Static Discharge) prevention procedures, and ensure that the 
maximum power specification for this channel (+27 dBm) is not exceeded.
Ch   2
      Freq
Display
Power
Ch   2
      Freq
      Pwr
dBm