Setting the Resolution and Measurement Rate
Operating Guide
2-15
2
Resolution Setting Example
For the following example, use the 10 MHz output from the reference 
timebase as the input to Channel 1.
1
Press the Resol key to enter the resolution-setting mode.
The current resolution setting is displayed (the current value and the 
indicator between the arrow keys are flashing to indicate that you can use 
the up- and down-arrow keys to change the setting).
2
Press the up-arrow or down-arrow key to decrease or increase 
the resolution.
 
You can press these keys as many times as necessary to locate the setting 
you want to use. The available resolution settings are 1 Hz (the default 
setting), 10 Hz, 100 Hz, 1 KHz, 10 KHz, 100 KHz, and 1 MHz.
3
Press  the  Enter key to activate your setting and exit the 
resolution-setting mode.
The setting you chose is now in effect. The number of digits displayed for 
the measurement is adjusted accordingly; you can observe the affect on 
measurement speed by monitoring the flash rate of the Gate indicator.
NOTE
The measurement resolution has a direct effect on the amount of time the 
Counter requires to complete a measurement. Measurements made at the 
Counter’s maximum resolution setting (1 Hz) are noticeably slower than 
at lower resolutions. This is especially noticeable when the Counter is set 
to average a number of measurements.